A chlorophyll online analyzer uses optical fluorescence to measure algal biomass, and its sensor and sampling lines are sensitive to freezing. Most units are rated for operation between 0 and 50 °C, and once ambient temperatures drop below freezing, residual water inside the tubing or measurement cell can expand and cause permanent damage. Antifreeze is therefore a targeted protective measure, not a routine additive.
The first and most basic trigger is sustained sub-zero ambient temperature. When winter conditions keep the air below 0 °C, or when the instrument is installed in northern alpine regions or exposed open-water sites where cold winds accelerate heat loss, ice formation becomes a real risk.
A practical threshold is a monitored water temperature of 5 °C or lower, at which point a low-temperature response should be activated. If the analyzer is housed indoors or in a temperature-controlled enclosure that stays within its rated operating range, antifreeze is unnecessary.
The second scenario is extended shutdown. If the analyzer is taken out of service for the winter, the sampling lines and measurement cell must first be flushed with clean water, then drained and dried with compressed air. A suitable volume of antifreeze is then injected into the lines, and both ends are sealed to prevent residual moisture from freezing and cracking the tubing or blocking flow paths.
Selection and concentration matter as much as timing. Food-grade antifreeze should be used, and the ethylene glycol concentration should not exceed 30%, because higher concentrations can alter the optical properties of the water sample and interfere with chlorophyll fluorescence readings.
Antifreeze must never enter the optical measurement path or the sample stream during normal operation; it is a protective fluid for idle lines, not a component of the analytical system. Before the instrument is restarted, the lines must be thoroughly flushed and the sensor recalibrated to prevent residual antifreeze from biasing subsequent measurements.

