Accurate measurement of chlorophyll-a concentration is crucial for assessing phytoplankton biomass and water quality in aquatic environments. Proper operation of laboratory chlorophyll analyzers, typically fluorometers or spectrophotometers, is fundamental to obtaining reliable data. The following key requirements must be observed.
1. Sample Preparation and Handling
Water samples must be collected using appropriate protocols to ensure representativeness. Immediately after collection, samples should be stored in dark, cool conditions and processed as soon as possible, ideally within 24 hours, to prevent pigment degradation. Filtration onto specific glass fiber filters (e.g., Whatman GF/F) under low vacuum pressure is standard. The filters must then be placed in a designated solvent (commonly 90% acetone) and stored in darkness at 4°C for extraction, typically for 20-24 hours.
2. Instrument Calibration and Setup
Prior to analysis, the instrument must be calibrated according to the manufacturer's guidelines. This includes a zero calibration using the pure extraction solvent (e.g., acetone) to establish a baseline. For fluorometers, calibration with a known chlorophyll-a standard is essential to convert fluorescence readings into concentration units. Ensure the instrument is placed on a stable, vibration-free surface and allowed sufficient warm-up time.
3. Measurement Procedure
Avoid exposing samples or extracts to direct light. Before measurement, the extract must be centrifuged or carefully decanted to remove filter debris. The clarified supernatant is then transferred into a clean, optical-grade cuvette. Wipe the cuvette's external surfaces thoroughly to remove fingerprints or droplets. Insert the cuvette correctly into the instrument's sample chamber. Record measurements promptly, and analyze samples in duplicate or triplicate to ensure reproducibility.
4. Data Recording and Quality Control
Meticulously document all parameters: sample ID, volume filtered, extraction volume, dilution factors, and raw instrument readings. Maintain a regular schedule for verifying calibration. Include quality control measures such as analyzing blank filters (processed identically but without a sample) to correct for any background interference and periodically testing a standard solution to check instrument drift.
5. Cleaning and Maintenance
Immediately after use, rinse the cuvette thoroughly with the appropriate solvent followed by clean water, and allow it to dry. Follow the manufacturer's instructions for routine maintenance to keep optical surfaces clean and the instrument in optimal condition. Keep a detailed log of all maintenance activities.
6. Safety Considerations
Most extraction solvents are volatile, flammable, and toxic. All procedures involving solvents must be conducted under a fume hood while wearing appropriate personal protective equipment (PPE), including lab coats, gloves, and safety goggles.
Adherence to these standardized requirements minimizes operational errors, ensures personnel safety, and guarantees the production of high-quality, comparable chlorophyll-a data for scientific and monitoring purposes. Always consult the specific operational manual of your instrument for detailed protocols.